TAS 200 Series Performance Verification
Bench Test Instruments and Handheld Oscilloscopes Basic Service
9
Set Up
Following a 20 minute warm-up period, preset the oscilloscope to the settings
listed below.
Table 12: Oscilloscope initial settings
Control Setting
INTENSITY Visible Display
CURSOR ON/OFF Off
VERTICAL
POSITION Midrange
MODE CH1
VOLTS/DIV 10 mV
VARIABLE CAL
GND In
HORIZONTAL
SEC/DIV 10 ms
POSITION Midrange
MAG Off (Out)
X-Y Off (Out)
SWEEP VARIABLE CAL
TRIGGER
SLOPE Posit ive (push switch out)
LEVEL Midrange
MODE AUTO
SOURCE CH 1
COUPLING DC