Maintenance
TDS1000B and TDS2000B Series Oscilloscope Service Manual
6-- 9
Installation. To install the flip feet, refer to Figure 8--4, and follow these steps:
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Hold the flip foot with the smooth side facing down.
3. Gently squeeze the foot so that the hinge on one side fits into the slot, and
twist slightly to snap into the slot on the other side.
Remove each front-panel knob by firmly grasping the knob (with pliers if
necessary) and pulling it away from the front panel.
NOTE. Wrap the plier jaws with tape (such as electrical) to avoid marring the
knob surface.
Install front-panel knobs by aligning the keyed knob with the shaft and pressing
the knob onto the shaft.
Remove the power button by firmly grasping the button with pliers, and pull it
straight out of the cabinet.
NOTE. Wrap the plier jaws with tape (such as electrical) to avoid marring the
button surface.
Install the button by pressing it onto the power button shaft until it snaps into
place.
You will need a torque-limiting Torx T-15 screwdriver and pliers for this
procedure.
Removal. To remove the rear case, refer to Figure 8--4, and follow these steps:
1. Remove the power button using the procedure on page 6--9.
2. Place the oscilloscope face down on a soft surface such as an anti-static mat.
3. Remove the five screws located on the rear case. Tilt the handle to gain better
access to the top two screws.
4. Lift the rear case off of the oscilloscope.
Front-Panel Knobs
Power Button
Rear Case