Reference
A typical isola
tion plot for the P7700 Series probes using a flex circuit-based
solder tip is shown in the following figure. When the probe is used with P7700
Series solder tips, note that channel isolation performance is highly dependent
on probe tip attachment lead length. Good channel isolation requires keeping the
interconnect lead length for both signal and ground connections very short.
Figure 24: Typical channel isolation
36 P7700 Series TriMode Probes Technical Reference