Specications
Table 6: Typical electrical characteristics (cont.)
Characteristic Description
TekFlex solder-in tips P77BRWSR
System noise DSP (all Input modes)
P7720 <4.2 mV
rms
<11.0 m V
rms
P7716 <4.0 mV
rms
<11.0 m V
rms
P7713 <3.8 mV
rms
<10.2 mV
rms
P7708 <3.6 mV
rms
<8.9 mV
rms
1
Single-e
nded, unused side terminated to 50 ohms
2
Measured as the -3 dB bandwidth for an ambient temperature range of 18 °C to 28 °C (64.4 °F to 82.4 °F)
Table 7: Electromagnetic compatibility (EMC)
Item Description
Certications, standards, and
classications
European Union
EC Council EMC Directive 89/336/EEC, amended by 93/68/EEC
Emiss
ions, EN55011, Class A
Immunity, IEC61000–4–2 and IEC6100–4–3
Austr
alia
AS/NZS 2064
United States FCC, CFR Title 47, Part 15, Subpart B, Class A.
Russ
ia
GOST complianc e
Emi
ssions
EN55011, Class A
Ra
diated emissions
EN 55011, Class A
Class A, 30-230 MHz at 40 dBuV/m (quasi-peak) at 10 m, 230-1000 MHz at 47 dBuV
(q
uasi-peak) at 10 m
Class B, 30-230 MHz at 30 dBuV/m (quasi-peak) at 10 m, 230-1000 MHz at 37 dBuV/m
(quasi-peak) at 10 m
Immunity
Electrostatic discharge (ES D) IEC 61000-4-2
IEC 61326, Performance Criterion “B”
42 P7700 Series TriMode Probes Technical Reference

Table of Contents

Other manuals for Tektronix TriMode P7700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TriMode P7700 Series and is the answer not in the manual?

Tektronix TriMode P7700 Series Specifications

General IconGeneral
BrandTektronix
ModelTriMode P7700 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals