Installation
Line Termination
Your instrument uses passive loop-through analog and reference inputs. Accordingly, the loopthrough inputs must be
terminated externally. It is important that this external termination meets accuracy and return loss requirements. Terminations
for loop-throughs on SDI or reference connectors in end-of-link setups must be 75 and DC coupled (good return loss
extends to DC). An appropriate termination would be a 75 End of Line terminator, Tektronix part number 011-0163-XX.
Compatibility of BNC Center Pins
Most video equipment BNC connectors, whether 50 or 75 ,usea50 standard center pin. Some laboratory 75 BNC
connectors
use a smaller diameter center pin. The BNC connectors on the instrument are designed to work with the
50 standard (large diameter) center pins.
Do not use connectors or terminators with the smaller center pins. They could cause intermittent connections.
WFM4000 and WFM5000 Waveform Monitors User Manual 5