SPECIFICATIONSTROUBLESHOOTING
1716
Power supply
Battery Powered Mode:
6V (4 x AAA-size alkaline-battery)
AC Adaptor Powered Mode:
100-240 V~, 50-60 Hz, 400 Ma
(Can be supplied by AC adaptor model
UE08WCP-060100SPA only!)
Display mode
Blue LCD with White Backlight
V.A. = 128mm(L) x 50mm(W)
Measurement mode
Oscillographic testing mode
Measurement range
Pressure: 0-40kpa(0~300mmHg)
pulse value:(40-199)times/minute
Accuracy
Pressure:
5-40 within ±0.4 kPa (3 mm Hg)
0-45(out of 5-40) within ±0.7 kPa
(5 mm Hg); Pulse Value: ±5%
Working condition
Temperature:5-40
Relative Humidity ≤85%RH
Atmospheric Pressure: 86-106 kPa
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm-32cm
Weight
Approx.385g(Excluding the dry cells)
External dimensions
Attachment
Approx.120*160*69mm
4*AAA alkaline batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
IP Classification
IP22
WARNING: No modification of this equipment is allowed.
Temperature:-20-60
Relative Humidity 10%-93%RH
Atmospheric Pressure: 50-106 kPa
Software Version
V01
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low
batteries
Error
massage
Display is dim or
will not light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Show on
the display
Batteries are low.
E 1 shows
The cuff is not secure.
Refasten the cuff and then
measure again.
E 2 shows The cuff is very tight
Refasten the cuff and then
measure again.
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
The monitor detected
motion while measuring.
movement can affect the
measurement.Relax for a
moment and then
measure again.
Measure incorrectly.
Relax for a moment and
then measure again.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
E 10 or
E 11 shows
E 20 or
E 21 shows
Eexx,shows on
the display.
Device Classification
Battery Powered Mode: Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
This section includes a list of error messages and frequently asked
questions for problems you may encounter with your blood pressure
monitor. If the products not operating as you think it should, check here
before arranging for servicing.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TRANSTEK LS805-B and is the answer not in the manual?

TRANSTEK LS805-B Specifications

General IconGeneral
BrandTRANSTEK
ModelLS805-B
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals