EasyManuals Logo
Home>TSI Instruments>Monitor>EVERWATCH 8610

TSI Instruments EVERWATCH 8610 User Manual

TSI Instruments EVERWATCH 8610
33 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Technical Section
17
Symptom
Possible Cause
Corrective Action
The display does
not indicate zero
at zero flow.
Velocity sensor.
Calibration
required.
Check velocity sensor location. Make sure it matches
installation instructions. The sensor should directly
reference room air. If required, mount a fume hood vent
kit.
The EverWatch monitor must be calibrated when
installed or moved.
The display
fluctuates.
Calibration.
Room turbulence.
The span may be incorrectly adjusted.
Check and adjust the display averaging period.
Check the effects of the room air distribution system on
the EverWatch monitor.
1. Turn the hood off.
2. Measure the air velocity at hood face.
3. Measure the air velocity at the velocity sensor.
The terminal throw velocity of a diffuser at the hood face
should be less than 50% of the average hood face
velocity. The terminal throw velocity at the velocity
sensor should be less than 10 ft/min (0.04 m/s). For
example, if the hood is running at 100 ft/min the terminal
throw velocity at the hood face should be less than 50
ft/min. The lower the terminal flow velocity, the better. If
the terminal throw velocity of a diffuser exceeds the
values listed above, move or readjust the diffuser.
The indicated
face velocity and
the average face
velocity are not
the same.
Sensor is dirty.
Calibration is off.
Check cleanliness of the velocity sensor. Remove all
contaminants.
Check the velocity sensor for obstructions. Remove any
obstructions.
Calibrate the EverWatch monitor.
The analog
output doesn’t
display correctly.
Incorrect output
selected.
Check that correct output is selected. VOLt selects a 0
to 10V output. CUrr selects a 4 to 20 mA output.
"SEnS Err" is
indicated on the
display.
Velocity is too
high.
Possible bad
sensor or sensor
cable.
Velocity exceeded 1500 ft/min. Adjust fan speed or
balancing damper.
Check the sensor wiring and connections for continuity.
Replace the velocity sensor and/or sensor to monitor
cable and calibrate the EverWatch monitor.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments EVERWATCH 8610 and is the answer not in the manual?

TSI Instruments EVERWATCH 8610 Specifications

General IconGeneral
BrandTSI Instruments
ModelEVERWATCH 8610
CategoryMonitor
LanguageEnglish

Related product manuals