20 Chapter 2
8. Make certain the alcohol
cartridge is clean and
insert the alcohol
cartridge into the alcohol
fill capsule. Turn the
capsule clockwise until it
locks in place.
9. Set the alcohol fill capsule
down and wait at least
two minutes while the
alcohol wick inside the
alcohol cartridge soaks up
the maximum amount of
alcohol.
10. Remove the alcohol
cartridge from the capsule
and gently shake it to
allow excess alcohol to
drip back into the alcohol
fill capsule. Stop when
excess alcohol is no
longer dripping; it is not
necessary to wait until the
outside surface of the
alcohol cartridge is dry.
11. Insert the alcohol cartridge into the cartridge cavity of the
PortaCount Fit Tester. It should slide in with little effort; DO NOT
force it. As you approach full insertion, firmly twist the alcohol
cartridge clockwise until it locks into position.