22 Chapter 2
Supply Power to the PortaCount Fit Tester
Using the AC connector provided, plug the PortaCount Fit Tester into
an AC power source. Power on the instrument with the on/off button
on the front. It will light up to indicate that power is on. To power
the instrument off, press and hold the power button for 1 second. It
will take up to 10 seconds to power down after that.
C A U T I O N
Use only the AC adapter provided (TSI part number
800531) and DO NOT substitute another adapter as
this will cause damage to the PortaCount Fit
Tester.
DO NOT block the fan or exhaust port on the back of
the instrument when it is turned on. Open air flow to
the back is important to the accuracy of the
measurement.
When first powered on, the instrument begins a one-minute warm up
period. The LED on the power button will strobe during the warm up
period. The PortaCount Fit Tester will connect to FitPro™ Ultra
Software after the warm up period is complete. The following
indicators will appear.
Wi-Fi capability is
enabled
Wi-Fi capability is not
enabled
Factory service is
recommended
Replenish alcohol in
wick cartridge