Model UT60H: OPERATING MANUAL
39
MAINTENANCE(2)
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
To replace the Meter’s fuse:
l Press the POWER to turn the Meter off and remove
all connections from the terminals.
l Remove the fixed rubber feet and 5 screws from the
case bottom.
l Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
l Install ONLY replacement fuses with the identical
type and specification as follows and make sure the
fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type fuse, φ5x20mm
Fuse 2: 10A, 250V, fast type fuse, φ5x20mm
l Rejoin the case bottom and case top, and install the
5 screws and the rubber feet.
Replacement of the fuses is seldom required. Burning
of a fuse always results from improper operation.
( figure 13)
Screw
C. Replacing the Battery (see figure 13)