EasyManuals Logo
Home>UNI-T>Measuring Instruments>UTL8500 Series

UNI-T UTL8500 Series User Manual

UNI-T UTL8500 Series
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
UTL8500 Series User Manual
18
Figure 4-1-1-1 Setting Interface of CC Mode
Operation Steps:
1. In initial interface, select [Mode], and then press [CC] to enter the setting interface of
<Constant Current CC>.
2. Move the cursor to current setting by pulse knob, and press the Enter button to
change the current value (the value changes from yellow to white) by keyboard or pulse
knob. Press the Enter button again to confirm. At this time, the set current value
changes from white to yellow.
3. Other parameters can be modified in a similar way.
4. Press the ON button, the electronic load starts to load, and the indicator light below the
button lights up. If you need to stop the load, press the ON button again, and the
running indicator light goes out.
Figure 4-1-1-2 Relationship of Voltage and Current in CC Mode
4.1.2 Constant Voltage Test
In CV mode, the electronic load changes current consumption to keep the input voltage as
set value.
Constant Current Mode