4.2 Daily Maintain
1. Protect the indicator from strong sunlight to prolong the
using life
2. Good connection between load cell and indicator. Far from
away from strong electric eld, magnetic eld.
3. Power o the indicator when lightning
4. Power o the indicator rstly before unplugging
4.3 Restore default parameter
Enter to calibration, Set C07=1. Press” PRINT” then press”
TOTAL” to exit saving setting. All parameter will go back to default
Note : Do not restore default parameter if you are not professional
ERR3
During calibration, the
input signal is negative
1. Check the connection
2. Check load cell for problems
3. Recalibration if still wrong
change the PCB
ERR4
During calibration, the
signal is unstable.
A er the platform is stable, start
calibration
ERR5 Change PCB
ERR2
During calibration , the
weights is below than
Min. required weights
e calibration weights
Minimum is 10% of Max. cap. Rec-
ommend 60%-80% of Max. Cap.