5125 & 5127 Disassembly
Note: Reassembly is reverse of Dis-Assembly
Note: Whenever the scale is opened for repair purposes a Calibration must be performed.
1. Make sure that the scale is powered
down by observing the readout
screen. If the screen is empty, it is
off. If not, wait for approximately 45
seconds for the scale to shut down.
2. Use a #1 Phillips screwdriver to
unscrew the two screws, remove the
battery access door and the
batteries.
3. Flip the scale over to access the
bottom of the scale. Peel the gray
tape from each corner half way or
until the four nuts underneath are
visible.