27
5102 Disassembly
Note: Reassembly is reverse of Disassembly
Note: Whenever the scale is opened for repair purposes a Calibration must be performed.
1. On the wall mount readout, use a #1
Phillips screwdriver to remove the
two screws of the battery panel
cover
2. Remove the six Duracell batteries
3. Using a 1/8” Allen wrench, remove
the two screws on each side of the
readout
4. Remove the readout from the wall
bracket