43
3.4 Place the viscometer into the ViscoClock plus
Press the guidance plate (7) up against the handle plate (9) (see Fig. 3).
- The following applies to the use of Ubbelohde and Micro Ubbelohde viscometers:
Insert the viscometer sloped through the guidance plate (6), and then insert it vertically into the
centering (2) prism.
> The capillary tube of the viscometer is located underneath the IR light barriers.
Lower the guidance plate (7).
Check for correct seating of the silicone plug on the fill tube (21) (silicone plug 8D) and capillary
tube (22) (silicone plug 8S).
The viscometer is now seated.
Attach the silicone cap (10) to the ventilation tube at a 90° angle!
With a lateral load on the ventilation tube, the glass tube may break!
- The following applies to Micro Oswald viscometers:
Insert the Micro Ostwald viscometer into the stand (3) of the ViscoClock plus so that the fill tube is
inserted into the elongated hole of the guidance plate (6).
Lower the guidance plate (7).
With Micro Ostwald viscometers, the manual pump set "suction" VZ 6554 is required.
Fig. 3 Inserting the viscometer into the ViscoClock plus
6
3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem SI Analytics ViscoClock plus and is the answer not in the manual?

Xylem SI Analytics ViscoClock plus Specifications

General IconGeneral
BrandXylem
ModelSI Analytics ViscoClock plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals