Wiring
Transmitter
46 / 54
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
Status input Sin
Do not connect a signal source with electric voltage.
The status input is provided for use of voltage-free contacts with the following specifica-
tion:
Switching status Resistance
Closed < 200 Ω
Open > 100 kΩ
ROTAMASS
Sin+
Sin-
Fig.38: Status input connection
7.4.7 Connecting power supply and external devices
NOTICE
Risk of damage to the flow meter due to incorrect power supply
The specified power supply must be observed (see General Specifications).
The power-supply cable must be designed for the power supply used with a
minimum diameter of 0.5 mm.
2
1
1 Transmitter back cover
2 Power supply cable gland