Malfunction of operation
Quick Reference Instruction Manual
Troubleshooting
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
51 / 54
9 Troubleshooting
All error messages and error codes that may appear in operation are described in the
software instruction manual. Possible malfunctions that may occur during commissioning
are explained below and remedying them is explained. If you cannot remedy the malfunc-
tion using these explanations, contact the Yokogawa service center.
9.1 Malfunction of operation
Malfunction Possible causes Remedy
Display on transmitter
not functioning
Power supply disconnected
Ensure that the unit is connected to the power
supply, see [}46]
Connection to main board
faulty
Check cable connections between display
and main board and connect properly, if
necessary.
Settings cannot be
made via IR switches
Incorrect settings in write-pro-
tect menu item
Switch off write-protect menu item via digital
communication or hardware switch.
Field communicator is
not detected
HART or Modbus DD not in-
stalled on field communicator
Save HART or Modbus DD file on field com-
municator.
Field communicator not con-
nected
Connect field communicator with Rotamass
Total Insight, see software instruction manual.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Total Insight series and is the answer not in the manual?

YOKOGAWA Total Insight series Specifications

General IconGeneral
BrandYOKOGAWA
ModelTotal Insight series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals