Page #15 background imageLoading...
Page #15 background image
2CMC486004M0201 11 C13
Revision: A User Manual
Installation
3.2 Environmental Considerations
Ingress protection
To comply with the protection requirements the product must be mounted in pro-
tection class IP 51 enclosures, or better, according to IEC60259.
Mechanical environment
In accordance with the Measuring Directive (2014/32/EU), the product complies
with M1, which means that it can be operated in “...locations with vibration and
shocks of low significance, e.g. for instruments fastened to light suporting struc-
tures subject to negligible vibrations and shocks transmitted from local blasting
or pile-driving activities, slamming doors, etc.”
Electromagnetic environment
In accordance with the Measuring Directive (2014/32/EU), the product complies
with E2, which means that it can be operated “...in locations with electro magnetic
disturbances corresponding to those likely to be found in other industrial build-
ings.”
Climatic environment
In order to work properly the product should not be operated outside the specified
temperature range of -25°C – +70°C.
In order to work properly the product should not exposed to humidity exceeding
the specified 75% yearly average, 95% on 30 days/year.