6 SCC-C Sample Gas Cooler Operator’s Manual 42/23-55 EN Rev. 1
Chapter 1 Preparing the Installation
Normal Operation
Normal Operation The SCC-C sample gas cooler is designed for cooling the sample gas and for
separating and removing the condensate.
The SCC-C sample gas cooler must not be used in hazardous areas. The model of
sample gas cooler featuring stainless-steel heat exchangers is suitable for use with
flammable sample gases.
Requirements for the Installation Site
Short Gas Paths The sample gas cooler should be installed as close as possible to the gas analyzer.
Adequate Air
Circulation
Ensure adequate natural air circulation around the sample gas cooler. Avoid heat
accumulation.
Protection against
Adverse Environ-
mental Conditions
The sample gas cooler should be protected from
• cold,
• radiated heat, e.g. from the sun, kilns or boilers,
• temperature fluctuations,
• strong air movement,
• dust deposits and dust penetration,
• corrosive atmosphere,
• vibration.
Climatic Conditions Ambient temperature During operation: +10 to +50 °C
Storage and transport: –25 to +60 °C
Relative humidity ≤ 75 % year-round average, ≤ 95 % on 30 days
per year, occasional light condensation
permissible