164
Measurement and storage conditions
앫 Blood glucose tests may be performed at the limit of the permitted
range (between +6 °C and +10 °C or between +40 °C and +44 °C).
When the temperature is at the limit of the permitted range, the
thermometer symbol flashes in the display.
Do not use results obtained at the limits of the permitted
temperature range as a basis for making therapeutic deci-
sions. The results may be incorrect. Incorrect results can
cause the wrong therapeutic decision to be taken and so pro-
duce serious adverse health effects.
앫 Tests cannot be performed at temperatures below +6 °C and above
+44 °C.
In this event the following message is displayed:
앫 Keep the finger pricker at temperatures between -25 °C and
+70 °C.
AC_CP_LCM_Manual_EN-UK final.indd 164_CP_LCM_Manual_EN-UK final.indd 164 20.6.2007 11:04:4020.6.2007 11:04:40