199
Technical data
13
In addition it meets the electromagnetic
emissions requirements according to EN
61326. Its electromagnetic emission is thus
low. Interference from other electrically pow-
ered equipment is not to be anticipated.
Performance evaluation
The performance data for the Accu-Chek
Compact Plus system (Accu-Chek Compact
Plus meter together with Accu-Chek Compact
test strips) were determined using capillary
blood from patients with diabetes (method
comparison, accuracy), venous blood (repeat-
ability) and control solutions (reproducibility).
The system is calibrated with venous blood
containing various levels of glucose. The ref-
erence values are obtained using the hexoki-
nase method. The method comparison was
conducted by comparing the results from the
system with those of the hexokinase method
with deproteination using an automatic ana-
lyzer. The hexokinase method is traceable to
an NIST standard (traceability).
The Accu-Chek Compact Plus system complies with the requirements of
EN ISO 15197.