Symbols, error messages and troubleshooting
Farbe: P 249 – Prüfmittelnummer: 203
If this happens … it means:

The drum compartment cover is not
properly closed.
Close the drum compartment cover. It
must close with an audible CLICK.
The ambient temperature is low and
the batteries are almost empty.
Move to a place where the
temperature is between +10 and
+40 °C and wait for the temperature
of the meter to adjust to this
temperature. Insert batteries (new or
recharged).
Condensation has affected the
electronics.
Allow the meter to dry slowly.
The meter is defective.
Please contact the Accu-Chek
Customer Care Line.
man_06919570001_01_EnCan.indd 182 18.01.2013 18:39:03

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH