Chapter 1: Your New Meter
Understanding
Chapter 2: Control Testing
Example
Range
Low 33–
High 311–
Low 43–
High 321–
16
Chapter 2: Control Testing ACCU-CHEK Customer Care Service Center 1-800-858-8072
Notes:
Do not operate without the nose cover attached. For more
information, see
Chapter 5: Maintenance and Troubleshooting.
When applying control solution, do not flood the test strip. Only apply
a small drop to the end of the test strip.
Do not bend the test strip.
Use only ACCU-CHEK Compact control solution with your ACCU-CHEK
Compact system or your test results may be inaccurate.
Never try to reuse a test strip that has been released from the meter.
Do not open the test drum compartment when the motor is running.
After a result has been displayed for 60 seconds, the meter turns off
automatically.
If the display goes blank before the test strip is released, press
to release the strip and turn off the meter.
If you don’t press the SET button before running a control test, the
meter will store the result as a normal blood test. This will include it
with your average blood sugar results and other trend information.
(This could be confusing for you and your healthcare provider.)
Gradient box behind
“Notes” prints 20c,
6m, 1k to white.
Gradient box behind
“Notes” prints 20c,
6m, 1k to white.
361_26758_Manual_AC_Compact.qxp 6/27/05 11:35 AM Page 16

Other manuals for Accu-Chek Compact

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact and is the answer not in the manual?

Accu-Chek Compact Specifications

General IconGeneral
BrandAccu-Chek
ModelCompact
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals