EasyManuals Logo
Home>Accu-Chek>Blood Glucose Meter>Nano

Accu-Chek Nano User Manual

Accu-Chek Nano
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
84
Your blood glucose may be extremely
high or a meter or a test strip error
has occurred. If your test result
matches how you feel, contact your
healthcare professional immediately.
If your test result does not match
how you feel, repeat the test and
refer to Chapter 3, Unusual Test Results. If this test
result does not match how you feel, perform a
control test with the control solution and a new test
strip. If the control result is within the acceptable
range, review the proper testing procedure and
repeat your blood glucose test with a new test strip.
If the E‑3 code still appears for your blood glucose
test, your blood glucose result may be extremely
high and above the system’s reading range. Contact
your healthcare professional immediately. If the
control result is not within the acceptable range, see
Chapter 2, Understanding Out‑of‑Range Control Test
Results.
The test strip may be damaged.
Remove the test strip and reinsert it,
or replace it if damaged.

Table of Contents

Other manuals for Accu-Chek Nano

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Nano and is the answer not in the manual?

Accu-Chek Nano Specifications

General IconGeneral
BrandAccu-Chek
ModelNano
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals