7-34
Adjustments
Manual Adjustments
FSK Deviation (Option 1EP Only)
Description
This adjustment modifies the FM Multiplier and out-of-band Kv arrays used by the signal
generator synthesizer module. The purpose is to provide a more accurate FM sensitivity
calibration while operating with FSK modulation.
Required Test Equipment
Vector Signal Analyzer
Procedure
Figure 7-33. FSK Deviation Test Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they are displayed on the PC.