7-53
Adjustments
Agilent Service Support Software
Running the Service Support Software
Starting the Software
For MS Windows version 3.x;
1. Open the Program Manager window.
2. Open the Agilent Service Support program group.
3. Select the Agilent Service Support icon.
For MS Windows 95, 98, NT:
1. Select Start.
2. Select Agilent Service Software for PC’s.
3. Select Agilent Service Software.
1. Log in using the User Information dialog box.
a. In the User Name field, type: User
(Case is important.)
b. In the Password field, type: User
c. Click OK.
Identifying the DUT
After logging in, the DUT Selection dialog box is displayed:
1. Ensure that 8648 is selected in the Select an instrument family list.
2. In the Select model list, select the instrument model of the DUT to be adjusted or
tested.
3. In the Serial Number box, enter the complete serial number of the DUT.