8-98
Performance Tests
8648C Test Record
Table 8-47. FM Accuracy Performance Test - Part 1
Frequency
(MHz)
Deviation
(kHz)
Limits (kHz)
Uncertainty
(±Hz)Lower Measured Upper
0.25 5 4.820 5.180 100
0.25 10 9.670 10.330 200
100 10 9.670 10.330 100
100 5 4.820 5.180 50
250 5 4.820 5.180 50
250 10 9.670 10.330 100
375 10 9.670 10.330 100
375 5 4.820 5.180 50
500 5 4.820 5.180 50
500 10 9.670 10.330 100
756.25 10 9.670 10.330 100
756.25 5 4.820 5.180 50
1000 5 4.820 5.180 50
1000 10 9.670 10.330 100
Table 8-48. FM Accuracy Performance Test - Part 2
RF
Frequency
8648C
(MHz)
LO
Frequency
8663A
(MHz)
Input
Frequency
8902A
(MHz)
Deviation
(kHz)
Limits (kHz)
Uncertainty
(±Hz)
Lower Measured Upper
1500 1501.5 700 5 4.790 5.210 50
1500 1501.5 700 10 9.640 10.360 100
2000 2001.5 700 5 4.790 5.210 50
2000 2001.5 700 10 9.640 10.360 100
2500 2501.5 700 5 4.730 5.270 50
2500 2501.5 700 10 9.580 10.420 100
3200 2500 700 5 4.730 5.270 50
3200 2500 700 10 9.580 10.420 100

Table of Contents

Other manuals for Agilent Technologies 8648A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8648A and is the answer not in the manual?

Agilent Technologies 8648A Specifications

General IconGeneral
BrandAgilent Technologies
Model8648A
CategoryTest Equipment
LanguageEnglish

Related product manuals