40 U1241A/U1242A User’s and Service Guide
5 Performance Tests and Calibration
Input-A Terminal Test
This test determines if the input warning of the A current terminal test is
functioning properly. The meter sounds an alerting beep when the test
lead is inserted to the A input terminal but the rotary switch is not set to
the corresponding A location. The display indicates a flashing AErr
annunciator until the test lead is removed from the A input terminal. This
warning alert is not available in T1/T2 temperature measurements mode.
Input-mA Terminal Alert Test
This test determines if the detected input voltage level is within the
accepted boundaries when the
μ
A/mA
input terminal is connected. The
meter sounds an alerting beep when the
μ
A/mA
input terminal detects a
voltage level of higher than 1.6 V. The display indicates a flashing
CErr
annunciator until the test lead is removed from the
μ
A/mA
input terminal.
Calibration Process
1 Prior to performing the verification tests, see Test Considerations.
2 Perform the verification tests to characterize the meter, see
Performance Verification Tests.
3 Unsecure the meter for calibration, see Calibration Security.
4 Prior to performing the adjustments, see Adjustment Considerations.
5 Perform the adjustment procedures, see Calibration Adjustments.
6 Secure the meter against unauthorized calibration, see Exiting
Adjustment Mode. Ensure that the meter has quit the Adjustment Mode
and turned off.
7 Record the new security code and calibration count in the meter’s
maintenance records.
NOTE
The alerting beep will not be impacted even if the beep function is disabled.

Table of Contents

Other manuals for Agilent Technologies U1242A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1242A and is the answer not in the manual?

Agilent Technologies U1242A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1242A
CategoryMultimeter
LanguageEnglish

Related product manuals