Performance Tests and Calibration 5
U1241A/U1242A User’s and Service Guide 55
Exiting Adjustment Mode
1 Remove all shorting plugs and connectors from the meter.
2 Record the new Calibration Count, see Calibration Count.
3 Press and simultaneously to exit the Adjustment Mode.
Power off and on the meter to return to normal measurement mode
and secured.
Step Test Function Cal Range Input Cal Item
U1241A U1242A
Move the test lead from “μA.mA” and “COM” terminal to “A” and “COM” terminal
Caution: Connect the calibrator to the meter's A” and “COM” terminal before applying 10 A
10 Turn the rotary switch to
A
position
Open Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
oPEn
10 A 10 A 10.000 A
11 Press to go to
A 10 A0.5 A, 70 Hz 0.500 A
function 1 A, 70 Hz 1.000 A
10 A, 70 Hz 10.000 A
12 Turn the rotary switch to
T1 or
T1T2 position
Short Dual banana plug with
copper wires short
between two terminals
SHrt
100 mV 100 mV 100.00 mV
13 Press to go to
T1 function K-type 0 ºC 000.0 ºC

Table of Contents

Other manuals for Agilent Technologies U1242A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1242A and is the answer not in the manual?

Agilent Technologies U1242A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1242A
CategoryMultimeter
LanguageEnglish

Related product manuals