2. Warning: The equipment is not capable of withstanding the 500V
electrical strength test as defined in Clause 6.1 of UL 60079-7:2017
and 6.3.13 of UL 60079-11:2014. This must be taken into account
during installation.
3. The 1410S1 needs an external IP54 that meets the requirements of
UL 60079-0.
7.1.5 Canada
N6 Canada Division 2 with Intrinsically Safe Outputs
Certificate
80009647 (CSA)
Standards
CAN/CSA C22.2 No 60079-0:2019, CAN/CSA
C22.2 No 60079-7:2016, CAN/CSA C22.2 No
60079-11:2014, CAN/CSA C22.2 No.25:2014,
CAN/CSA C22.2 No.61010-1-12 3rd Edition
Markings with
Intrinsically Safe into
Division 1 or Zone 0
Class I Division 2, Groups A, B, C, D; Intrinsically
Safe outputs to Class I, Division 1, Groups A, B, C,
D; Class I, Zone 2 Ex ec [ia Ga] IIC T4 Gc
Markings with
Intrinsically Safe into
Division 2 or Zone 2
Class I Division 2, Groups A, B, C, D; Intrinsically
Safe outputs to Class I, Division 2, Groups A, B, C,
D; Class I, Zone 2 Ex ec [ic] IIC T4 Gc
Temperature code
T4 (-40 °C ≤ T
a
≤ 70 °C)
Special Conditions for Safe Use(X)
1. Warning: Potential electrostatic charging hazard - See instructions.
AVERTISSEMENT - RISQUE D'EXPLOSION. NE PAS DÉBRANCHER
PENDANT QUE LE CIRCUIT EST SOUS TENSION.
2. Warning: The equipment is not capable of withstanding the 500V
electrical strength test as defined in Clause 6.1 of CAN/CSA
60079-7:2016 and 6.3.13 of CAN/CSA 60079-11:2014. This must be
taken into account during installation.
Avertissement: L'équipement n'est pas capable de résister au test de
résistance électrique de 500V tel que défini dans la clause 6.1 de
CAN / CSA 60079-7:2016 et 6.3.13 de CAN / CSA 60079-11:2014.
Ceci doit être pris en compte lors de l'installation.
3. The 1410S1 needs an external IP54 that meets the requirements of
CAN/CSA 60079-0.
Le 1410S1 a besoin d'un IP54 externe qui répond aux exigences de
CAN/CSA 60079-0.
Quick Start Guide July 2022
30 Emerson.com