Stik Thermometer
Safety Information
3
If the Product overheats or is exposed to sudden physical shock, examine it for damage that
can cause a safety risk. If possible, compare the displayed temperature to a known reference
before Product use. If not sure, send the Product to Fluke Corporation. Refer to “How to
Contact Fluke”.
Substitution of components will impair suitability for hazardous locations.
ATEX Safety Information
This manual contains data and safety regulations that must be followed for safe, reliable
use of the Product in hazardous areas under the detailed conditions. If you do not follow
these instructions, personal injury or Product damage can occur. Violation of applicable
legislation can also occur. Read the full manual before you use the Product. To make sure
of safe Product operation, fully follow all instructions and warnings in the manual. If you
are not sure (due to translation and/or printing errors), refer to the English manual.
An “Ex-hazardous area,” as used in this manual, refers to an area made hazardous by the
possible presence of flammable or explosive vapors. These areas are also referred to as
“hazardous locations.”
Ex ib IIB T4 Gb (–10 °C
ITS10ATEX27114X
Ex ib IIB T4 Gb
II 2 G IECEx ITS10.0049 0344
Manufactured by Martel Electronics, Inc.,
3 Corporate Park Dr.
Derry, NH, USA
≤ Ta +50 °C)
gjo006.eps
Symbols
International symbols used on the Product and in this manual are explained in Table 1.
Table 1. Symbols
Symbol Meaning Symbol Meaning
J
Earth ground
M
Battery
W
Risk of Danger. Important information.
See Manual.
X
Hazardous Voltage
~
Do not dispose of this product as
unsorted municipal waste. Go to
Fluke’s website for recycling
information.
(
Conforms to ATEX requirements
P
Conforms to relevant European Union directives.

Other manuals for Fluke 1551A Ex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1551A Ex and is the answer not in the manual?

Fluke 1551A Ex Specifications

1551A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−50 °C to 160 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1551A-94.8 mm x 229 mm

1552A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−80 °C to 300 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1552A6.35 mm x 305 mm