Stik Thermometer
Calibration
7
Record as-found Calibration Parameters
Before any adjustments are made, the present values of the Product calibration
parameters should be recorded. Connect the serial cable between the computer and the
Product. The commands used to read the values of the calibration parameters are given in
Table 3. Commands must be terminated with a carriage return character. Table 3 may
also be used to record the values of the calibration parameters.
Table 3. As-found Calibration Parameters
Parameter Command Value (fill in)
TEMP1 CAL:USER:TEMP1?
ADJ1 CAL:USER:ADJ1?
TEMP2 CAL:USER:TEMP2?
ADJ2 CAL:USER:ADJ2?
TEMP3 CAL:USER:TEMP3?
ADJ3 CAL:USER:ADJ3?
DATE CAL:DEV:DATE?
Evaluate as-found Measurement Errors
The Product should be tested to evaluate its accuracy before any adjustments are made. It
is recommended that it be tested near the five temperatures listed in Table 4, depending
on the model.
Table 4. Test Temperatures
Point 1551A Ex 1552A Ex
1 -50 °C -80 °C
2 -25 °C -25 °C
3 0 °C 0 °C
4 100 °C 100 °C
5 157 °C 300 °C
At each of the test temperatures, readings are recorded of the temperature of the reference
thermometer and the indication of the Product. Before recording the readings, make sure
the temperatures of the bath, reference thermometer, and 1551A Ex/1552A Ex
Thermometer are settled and stable.

Other manuals for Fluke 1551A Ex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1551A Ex and is the answer not in the manual?

Fluke 1551A Ex Specifications

1551A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−50 °C to 160 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1551A-94.8 mm x 229 mm

1552A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−80 °C to 300 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1552A6.35 mm x 305 mm