1551A Ex/1552A Ex
Calibration Manual
6
Recommended Equipment
The equipment listed in Table 2 will satisfy the calibration requirements:
Table 2. Recommended Equipment
Equipment Recommended Model
Temperature Baths Fluke 6331 40 °C to 300 °C Hot Bath
Fluke 7381 -80 °C to 110 °C Cold Bath
Fluke 7196 Equilibration Block
Reference Thermometer Fluke 5699-S Extended Range Metal-Sheath SPRT
Reference Thermometer Readout 1594A Super-Thermometer readout
Computer
Computer with an RS-232 com port
RS-232 serial cable
Terminal software to communicate with the 1551A
Ex/1552A Ex Thermometer
Initial Setup
Check the batteries in the Product and replace them if necessary. The battery level
indication is found in the Setup menu.
Several settings in the Setup menu relate to measurement and should be set as follows:
Sample Rate: 1 per second
Damping: OFF
Display Resolution: 0.001
Auto Power Off: OFF
The unit of measurement in the 1551A Ex/1552A Ex Thermometer may be set to °C or
°F, but must match the unit of the reference thermometer.
The baud rate in the computer software must match the baud rate of the 1551A Ex/1552A
Ex Thermometer, which is usually 9600 but may be set to 2400.

Other manuals for Fluke 1551A Ex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1551A Ex and is the answer not in the manual?

Fluke 1551A Ex Specifications

1551A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−50 °C to 160 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1551A-94.8 mm x 229 mm

1552A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−80 °C to 300 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1552A6.35 mm x 305 mm