787
Users Manual
36
Specifications
All specifications apply from +18°C to +28°C unless
stated otherwise.
All specifications assume a 5 minute warmup period.
The standard specification interval is 1 year.
Note
“Counts” means number of increments or
decrements of the least significant digit.
DC Volts Measurement
Range (V dc) Resolution Accuracy, ±(% of Reading + Counts)
4.000 0.001V 0.1% + 1
40.00 0.01V 0.1% + 1
400.0 0.1V 0.1% + 1
1000 1V 0.1% + 1
Input impedance: 10 M
(nominal), < 100 pF
Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz
Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz
Overvoltage protection: 1000V

Other manuals for Fluke ProcessMeter 787

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ProcessMeter 787 and is the answer not in the manual?

Fluke ProcessMeter 787 Specifications

General IconGeneral
BrandFluke
ModelProcessMeter 787
CategoryMultimeter
LanguageEnglish