© 2007 by General Electric Company. All rights reserved. M1110150 006 29
Chapter 8: Specications
NOTE: All specications are nominal and are subject to change without notice.
Electrical Specications
Input 1.5A @ 100 – 240 V~, 50/60 Hz
Fuses T3.15A @ 250V~, slo-blo type (qty. 2)
Leakage Current < 300 μA @ 264 V~
Ground impedance < 0.1 ohm from ground pin of the power inlet module
to any exposed metal surface.
Environmental Operating Conditions
Ambient temperature +10°C to +35°C (50 to 95°F).
Humidity 10% to 90% RH non-condensing
Atmospheric pressure 70 kPa to 106 kPa
Storage Requirements
Temperature -40°C to +70°C (-40 to 158°F).
Humidity 0% to 100% RH non-condensing
Atmospheric pressure 50 kPa to 106 kPa