IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-8 - System Configuration 3 - 33
3-8-2-1 Enter Location, Adjust Date/Time, Select Language
NOTE: In R4 and later, the user can set up user ID restricting the access to Utility, including restricting
access to service. Please refer to the user manual for more information.
The System/General screen allows you to specify hospital name and system date and time.
1.) Select Utility on the touch panel.
2.) Select System and then select General.
Figure 3-23 System/General Preset Menu
3.) Set Hospital name, Department name, Language, Units, Time/Date format as needed.
4.) To save the changes, select Save on the bottom-left. Select Exit to return to scanning. In some
cases, you may need to reboot the system for the change to take effect.
Table 3-1 Location
Preset Parameter Description
Hospital Type the institution’s name.
Department Type the institution's department name.
Preset Region (restart
needed)
Select region (None, Americas, Asia, Europe or Japan).
Language (restart needed) Select the appropriate language from the drop-down list.
Note: If you select Japanese (JPN), only the warning and status messages are
displayed in Japanese. You can not type in Japanese.
Units Select metric or US units of measurement.
Regional Options Select to set up the keyboard.