GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3 - 74 Section 3-12 - Configuring Connectivity
3-12-5 TCPIP
This configuration category enables users with administrative rights to set the TCPIP for the system and
connected remote archive.
1.) Type the name of the Ultrasound system in the Computer Name field.
2.) In the IP settings section, identify the ultrasound system to the rest of the network by one of the
following:
- DO NOT enable DHCP.
- Type the IP-Address (acquire unique static IP address from hospital network administrator),
Subnet Mask, and Default Gateway (if applicable).
NOTE: Do not set up the system with DHCP. The IP address MUST BE static for the diagnostic and DICOM to
function correctly.
3.) Select Save settings.
4.) Re-boot the ultrasound system.
NOTE: TCPIP settings do not get restored when restoring backups. This is per system design. The LOGIQ
S8 IP address MUST BE unique.
Figure 3-47 Connectivity TCPIP Preset Menu
Table 3-9 Computer Name
Preset Parameter Description
Computer Name Type the unique name for the Ultrasound system (no spaces in name).