Section 9-8 - Electrical Safety Tests 9 - 25
9-8-7-2 Generic Procedure (cont’d)
For each combination, the operating controls, such as the lead switch, should be operated to find the
worst case condition.
Figure 9-8 Test Circuit for Measuring Patient Lead Leakage