Care and Maintenance
10-6 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
10-4 Maintenance task schedule
10-4-1 How often should maintenance tasks be performed?
The Care and Maintenance task schedule (provided in
Table 10-1 on page 10-6) specifies how often your LOGIQ V3/
V5/V5 Expert should be serviced and outlines items requiring
special attention.
NOTE: It is the customer’s responsibility to ensure the LOGIQ V3/V5/V5
Expert care and maintenance is performed as scheduled in
order to retain its high level of safety, dependability and
performance.
Your GE Service Representative has an in-depth knowledge of
your LOGIQ V3/V5/V5 Expert and can best provide competent,
efficient service. Contact GE for coverage information and/or
price for service.
The service procedures and recommended intervals shown in
the Care and Maintenance Task Schedule assumes that you
use your LOGIQ V3/V5/V5 Expert for an average patient load
(10-12 per day) and not use it as a primary mobile Ultrasound
system which is transported between diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load,
then it is recommended to increase the periodic maintenance
frequencies.
Table 10-1: Customer Care Schedule
Service at Indicated
Time Daily Weekly Monthly
Per
Facilities
QA
Program
Notes
Clean Probes •* * or before each use
Inspect AC Mains Cable • Mobile Ultrasound
system: Check
Weekly
Inspect Cables and
Connectors
•
Clean Console •