9-10 MAC 1200 resting ECG analysis system Revision D
2012250-095
Specifications: Basic Information
Mains Operation
Instrument design in protection class I according to IEC 601-1
Rated voltage range: 95 to 240 V
Operating voltage range: 85 to 264 V,
49 to 65 Hz
Rated current: 0.2 ... 0.6 A
Fuse: 2 x T 1.25 A, 5 x 20
Typical power consumption:
Battery charging: 14 W
Maximum power consumption: 29 W
Battery Operation
Type: nickel-cadmium
Rated battery voltage: 18 V
Rated battery capacity: 1.3 Ah
Fully charged battery sufficient for up to 50 12 Lead Mode, 1-page ECGs, if
unit is only switched on to record the ECGs
Battery charge time: approximately 4 hours
(Minimum charge time for one 12 Lead Mode
ECG: 10 minutes)
Battery life: approximately 2 to 3 years, replacement by service only
Lithium battery for built-in clock, battery life: approximately 5 years,
replacement by service only
Operational Readiness
After successful self-test, approximately 20 seconds after power-up
Operating Position
Horizontal

Table of Contents

Other manuals for GE MAC 1200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 and is the answer not in the manual?

GE MAC 1200 Specifications

General IconGeneral
BrandGE
ModelMAC 1200
CategoryMedical Equipment
LanguageEnglish

Related product manuals