3-20 MAC 1200 resting ECG analysis system Revision D
2012250-095
System Test Function: Serial Number
Serial Number
The serial number of the device is displayed by pressing the N key.
The serial number is only displayed and cannot be changed.
The serial number is entered during the manufacturing process.
After replacing the PCB Control CS_CI, during the power up sequence, the
serial number of the device has to be entered.
Terminate with the X key.