Appendix B: Specifications
Optima XR220amx X-Ray System 54400222-1EN Rev 1 DRAFT 11092010 B-4
Operator Manual © 2010 General Electric Company. All rights reserved.
User Interface
x Integrated 15 in (38.1 cm) touch-screen user interface
x Monitor matrix size 1024 x 768 pixels
x 250GB or greater for storage of more than ~30,000 uncompressed images
x Images can be stored on a CD with the included external CD drive
x USB outlets for external drive, barcode reader, service mouse and keyboard
Networking and DICOM
Worklist can be retrieved from HIS/RIS systems and images can be transmitted through the DICOM inter-
face to printers, archival devices (PACS), servers or review workstations.
x RJ45 10/100/1000 Base T Ethernet port or optional 802.11 a/b/g/n wireless connection
x IHE and DICOM 3.0 compliant services:
– DICOM IOD generated – DX
– Image auto send/transfer –DX or CR (support multiple destinations)
– DICOM storage – SCU
– DICOM storage commitment SCP, when Fixed IP address is provided (programmable auto-delete
function)
– DICOM modality worklist for HIS/RIS – SCU (with programmable auto refresh)
– DICOM grayscale print – manual and auto (with print layout options)
– Multi-format printing
– DICOM CD-R media exchange (DX only)
– Verification of connectivity for SCU & SCP services
– DICOM Modality Perform Procedure Step (MPPS) feedback to the HIS/RIS (SPS PPS) (configurable)
– VOI LUT burn on send (configurable)
Please refer to DICOM conformance statement for complete definition of supported DICOM services
Drive
x Battery operated
x Self propelled single drive handle control
x Variable speed of up to 5 km/h (3.1 mph) forward and reverse, automatically adjusts to the operator’s
pace; variable to adjust to the operator’s pace.
x Capable of climbing an incline of 7°
DRAFT REVIEW COPY