Appendix B: Specifications
Optima XR220amx X-Ray System 54400222-1EN Rev 1 DRAFT 11092010 B-6
Operator Manual © 2010 General Electric Company. All rights reserved.
Environmental Conditions
System
Digital Detector
Temperature
Operating +10° C to +35° C
Non-Operating System (except
Detector)
-10° C to +45° C
Humidity
Operating 30% to 75% RH, non-condensing (maximum change 30% per hour)
Non-Operating 25% to 85% RH, non-condensing (maximum change 30% per hour)
Altitude
Operating -30 m to +3,000 m relative to sea level
Non-Operating -30 m to +3,000 m relative to sea level
Atmosphere Pressure
Operating 106 kPa to 70.1 kPa
Non-Operating 106 kPa to 70 kPa
Audible Noise
Operating 60 dBA maximum audible noise at any point 1 meter from system
Idle 50 dBA maximum audible noise at any point 1 meter from system
Temperature
Operating +10° C to +35° C
Non-Operating 0° C to 50° C (maximum change 50° C per hour)
Humidity
Operating 10% to 80% RH, non-condensing (maximum change 30% per hour)
Non-Operating 10% to 95% RH, non-condensing (maximum change 30% per hour)
DRAFT REVIEW COPY
CTRL+SHIFT+F to search using Search Index (faster)

Table of Contents

Other manuals for GE Optima XR200amx

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Optima XR200amx and is the answer not in the manual?

GE Optima XR200amx Specifications

General IconGeneral
BrandGE
ModelOptima XR200amx
CategoryMedical Equipment
LanguageEnglish

Related product manuals