Operating Instructions
12 VeriDri™ User’s Manual
6.3 Evaluating the Probe
To evaluate the probe, complete the following steps:
1. Re-connect the cable to the transmitter module and measure the
ambient dew point. Make sure you measure the same ambient air as
measured in step 2 on page 10.
2. Compare the two ambient air readings. If the new ambient air
reading is within ±2°C (±3.6°F) of the first reading, the cleaned
probe is properly calibrated and may be reinstalled. If the two
readings not within that tolerance, proceed to step 3 below.
3. If the probe is still not reading the ambient air accurately, repeat the
cleaning procedure using soaking times that are 5 times the
previous cleaning sequence, until two consecutive ambient air
readings are within ±2°C (±3.6°F).
If the above cleaning procedure does not result in accurate readings, contact
GE for assistance.