strona 14 INSTRUKCJA SERWISOWA GHC201.07 i GHC201.08
PPUH „GECO” Sp. z o.o. Wydanie I OD DNIA 2018-02-22
the value of parameters ‘d0’ and ‘d1’ (d0<d1)
4. Compressor does
not switch on although
it is connected
Check:
- if there is voltage 230V on clamps P1 or P2
- value of parameters ‘r1’ or ‘r2’
5. Heater does not
switch on although it is
connected
Check:
- if there is voltage 230V on clamps P1 or P2
- value of parameters ‘r1’ or ‘r2’
6. Buzzer does not
switch on although it is
connected
Check:
- if there is voltage 230V on clamps P1 or P2
- value of parameters ‘r1’ or ‘r2’
- value of ‘d8’ parameter
7. Dots on the display
flash, no possibility to
switch it on
Check :
- the value of feeding voltage
- condition of feeding connectors
- whether feeding connectors are tightened up
8. ‘’abnormal’
‘strange’ operation of
the unit
Check:
- if there is voltage 230V on feeding clamps L and N
- condition of feeding connectors
- ‘zeroing’ of cooling unit
- condition of electric system and the number of units connected to one
phase (sticker on a case)
- weather GHC201.08 was not wetted by water or other liquid
- if GHC201.08 is not exposed to humidity or sudden changes of
temperature
9. The appliance does
not reach the set
temperature and does
not cool
Check:
- what temperature was programmed by the user
- set values of particular parameters, in particular ‘c2’, ‘c5’, ‘d0’, ‘d1’
- the method and place of fixing of the temperature sensor
- WHETHER THE UNIT STANDS IN A DRAFT OR IN SUNLIGHT!!!
- whether ventilators or air conditioning are installed on the ceiling or
nearby
IX. RETURNS FOR REPAIR
PPUH ‘GECO’ reserves the right to refuse a free repair of the unit, if
the seals are broken !!!
P.P.U.H. ‘Geco’ Sp. z o. o. is not responsible for loses and damages resulting from
provision of information on the method of making changes in the system data of
GHC201.07 or GHC201.08 by the producer or its service to the final client, incorrect or
non-professional assembly and for loses caused by defective operation of the
appliance.