2.4.2 Disinfectant (D2)
Let the parts soak in disinfectant.
Pre-disinfectant
• Aniosyme DD1
• Gigazyme Plus
• Salvanios pH7
Disinfectant
• Alcohol (ethyl or isopropyl alcohol 70%)
• Anioxyde 1000
• Cidex OPA
• Actanios HLD (prion decontamination
agent), may cause discoloration
Important: Follow the disinfectant
manufacturer’s recommendations and
instructions, otherwise the cassette may be
damaged.
2.4.3 Rinse after disinfection (D2)
• Rinse the parts thoroughly in water to
remove all traces of disinfectant. Let the
water flow through the parts.
• Carefully shake and tilt the cassette, turn
upside down and repeat.
• Repeat these steps at least 5-7 times.
Important: Mineral deposits from
disinfectant on the expiratory cassette
affect the function. It is important to rinse
the expiratory cassette thoroughly.
Residues from chemicals can affect the
patient, cause leakage and extra stress on
the material.