2.4.4 Drying alternatives
D2D1
Dismantle
Wipe off/discard
Rinse
Disinfectant
Rinse
Washer-
Disinfector
Dry
Assemble
The expiratory cassette must be dried before
use (if not dry, the expiratory cassette may not
pass the Pre-use check).
There are several drying alternatives for the
expiratory cassette:
Drying cabinet 1 hour in maximum 70°C
(158°F); if available, connect a 22 mm air
hose to the expiratory cassette for extra
drying effect, or
Drying the cassette in room air 12-24 hours,
depending on surrounding conditions.
Recommended position in a drying cabinet
Recommended position in room air
If the Pre-use check is not passed after drying
then:
Run the cassette in a ventilator system with
a test lung for 10 minutes, or
Dry the cassette in room air (if no drying
cabinet is available) 12-24 hours, depending
on surrounding conditions.
CAUTION: Never dry the cassette by
applying high-pressure air as the internal
tubing may be damaged.
Note: Drying may not be necessary if the
washer-disinfector has a drying phase.
SERVO-i/s, User's Manual
15
| Routine cleaning |
2 |

Other manuals for Getinge Maquet SERVO-i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-i and is the answer not in the manual?

Getinge Maquet SERVO-i Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-i
CategoryMedical Equipment
LanguageEnglish

Related product manuals