SCPI Programming Reference
HDG2000 Series Arbitrary Waveform Generator 80
SOURce<n>:PULSe:TRANsition:TRAiling? [MINimum|MAXimum]
Sets the pulse edge time on the leading, trailing, or both edges of a pulse.
Set leading edge time on CH1 to 10 ns: SOURce1:FUNCtion:PULSe:TRANsition:LEADing
0.00000001
Set trailing edge time on CH1 to 10 ns: SOURce1:FUNCtion:PULSe:TRANsition:TRAiling
0.00000001
The leading edge time applies to rising edge, and represents the time from the 10% threshold
to the 90% threshold of the edge; the trailing edge represents the time from the 90% threshold
to the 10% threshold.
The specified edge time must fit within the specified pulse width and period. The instrument will
limit the edge time to accommodate the specified pulse width or duty cycle.
5.2.13 HARMonicSubsystem
SOURce<n>:HARMonic:AMPL –Harmonic amplitude
SOURce<n>:HARMonic:PHASe –Harmonic phase
SOURce<n>:HARMonic:ORDEr –Harmonic order
SOURce<n>:HARMonic:TYPe –Harmonic type
SOURce<n>:HARMonic:AMPL <sn>,<value>|MINimum|MAXimum
SOURce<n>:HARMonic:AMPL?
Set Harmonic amplitude
<sn> integer, 2 to 16, default 2;
<value>Continuous real Numbers
Sets the amplitude of the second harmonic to 1Vpp on CH1: SOURce1:HARMonic:AMPL 2,1
SOURce<n>:HARMonic:PHASe <sn>,<value>|MINimum|MAXimum
SOURce<n>:HARMonic:PHASe? <sn>[,MINimum|MAXimum]
Set Harmonic phase.
<sn> integer, 2 to 16, default 2;
<value>Continuous real Numbers, 0 to
360degrees, default 0
Set the second Harmonic phase on CH1 to 90 degrees: SOURce1:HARMonic:PHASe 2,90