Handheld Thermography Camera User Manual
13
Chapter 4 Temperature Measurement
The thermography (temperature measurement) function provides the real-time temperature of
the scene and display it on the left of your screen. The thermography function is turned on by
default.
4.1 Set Thermography Parameters
You can set thermography parameters to improve the accuracy of temperature measurement.
Steps
1. Go to Local Settings Thermography Settings.
2. Set the Thermography Range, and Emissivity, etc.
Thermography Range
Select the temperature measurement range. The device can detect the temperature and
switch thermography range automatically in Auto Switch mode.
Emissivity
Refer to Common Material Emissivity Reference to set the emissivity of your target.
Reflection Temperature
If any object (not the target) of high temperature is in the scene, and the target emissivity is
low, set the reflection temperature as the high temperature to correct the thermography
effect.
Distance
The distance between the target and the device. You can customize the target distance or
select the target distance as Near, Middle, or Far.
Environment Temperature
The average temperature of the environment.
Humidity
Set the relative humidity of current environment.
Note
Go to Local Settings Image Settings Display Settings to enable/disable the emissivity
display.
3. Return to previous menu to save the settings.

Other manuals for Hikmicro M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hikmicro M Series and is the answer not in the manual?

Hikmicro M Series Specifications

General IconGeneral
BrandHikmicro
ModelM Series
CategoryThermal cameras
LanguageEnglish

Related product manuals