AlcoQuant 6020 plus
Doc no.: 059-07-1001779_A6020plus_WL_GB-5.2.1 / 12.2021 35
7 Technical specifications
Measurement mode : active and passive
Measuring range : 0 to 5.5 ‰ BAC*
Measuring accuracy : Max. error tolerance in temperature range from -
5°C to 45°C
• Range 0 to 1 ‰
:
± 0.05 ‰ absolute
• Range > 1 to 2 ‰
:
± 5% of measurement value
• Range > 2 ‰
:
± 10% of measurement value
Reproducibility with ethanol
standard in temperature range from -5°C to
45°C
• Range 0 to 1 ‰
0,017 ‰
• Range ≥ 1 ‰
1,7 % of measurement volume, the higher
volume is valid
Ambient conditions
• Operating temperature range
:
-10°C to +50°C
• Storage temperature range
:
-25°C to +70°C
• Storage temperature (sensor)
:
0°C to +25°C, optimal
• Ambient pressure
:
600 to 1400 hPa
• Ambient humidity
:
20 to 98% rel. humidity
• Minimum exhalation volume
:
1.2 litres
Batteries or accumulator (optional) : 4x LR6 (AA) alkaline Mignon batteries
(Industrial by Duracell recommended) or
Honeywell Healthcare Solutions GmbH -Akku
NiMH 4.8V (4 x 2100 mAh)
Charging station (optional) : protection class II
Input: AC 100–240 V/50–60 Hz/93 mA
Output: DC 5.6 V/500 mA/2.8 VA
Charging time : approx. 4 hours
Measurements per battery
pack/accumulator charging
: approx. 2000 measurements, depending on the
batteries and the ambient conditions
Device interface : USB and RS-232 interface
Data can be transmitted to a PC.
Part no. 1002249 and 1002251: Wireless
interface (IEEE 802.15.1) for usage with
wireless printer (part no. 1002255).